BioChrom Labs, Inc.
BioChrom Labs, Inc.
BioChrom Labs, Inc.
BioChrom Labs, Inc.
HomeWhat's NewHYDROCELL HPLC ColumnBiocompatible-CoatingCareer OpportunitiesDistributor InformationOrdering InformationContact Us

HYDROCELL HPLC Column
Ion Exchange HPLC Columns
Hydrophobic Interaction HPLC Columns
Non-Porous High Speed Ion Exchange Columns
Non-porous High Speed Hydrophobic Interaction Columns
Nucleic Acid HPLC Columns
Reversed Phase HPLC Columns
Size Exclusion HPLC Columns
Bulk Media
Guard Columns
Accessories
Product Selection Guide
Trouble Shooting
 

HYDROCELL Trouble Shooting


CONTENTS:

  1. Column Cleaning Protocol (Baseline drift and spike peaks)
  2. No Retention of Biomolecules in HYDROCELL Columns
  3. High Back Pressure

1. HYDROCELL Column Cleaning Protocol

If the peaks have broadened and the baseline has drifted when you use a HYDROCELL Ion Exchange or Hydrophobic Interaction Column, the column needs to be reactivated and regenerated. The procedure for reactivating and regenerating a 4.6 mm HYDROCELL column is as follows:

1. Flush with 80% methanol in water at a flow rate of 3 mL / minute for 10
minutes.

2. Flush with water, and then 20% methanol in 1 M NaCl solution at a flow
rate of 3 mL / minute for 10 minutes.

3. Flush with B Buffer at a flow rate of 3 mL / minute for 10 minutes.

4. Equilibrate with the starting buffer at a flow rate of 2 mL / minute.

5. Run a 10 minute blank gradient from 0 - 100 % B buffer at a flow rate of
2 mL / minute.

Repeat this process, if necessary, until the baseline has stabilized. Adjust the flow rates and volumes for other I.D. columns proportionately.

Alternative Procedure: Elute column with 1.0 M NaOH aqueous solution continuously, at a flow rate of 1 mL / minute for 1 hour or leave 1.0 M NaOH inside the column overnight.

Back To Top


2. No Retention of Biomolecules in HYDROCELL Ion Exchange or Hydrophobic Interaction Columns

HYDROCELL columns with the proprietary hydrophilic oligomers coating are very hydrophilic and do not have non-specific adsorption and binding. These columns do not need detergents, surfactants or organic solvents in the buffers to circumvent the non-specific binding that might occur in the silica-based or other polymeric packings in the separation of biomolecules. If you have problems of retention of your samples in HYDROCELL columns, please check the preparation of your samples, buffers, and the equilibrium time with the first buffer before injection of your samples.

(A) Sample Preparation:

1. Solid samples: The samples are prepared by dissolving about 3 mg of samples in 1 ml of the first buffer. It is necessary, likewise, in hydrophobic interactive chromatography, that a small volume of secondary buffer can be added to increase the solubility of the samples.

2. Liquid samples: The liquid samples should have detergents, surfactants, and solvents removed that will affect the retention of the samples in the separation, and adjust their pH value and salt concentration as close as possible to the first buffer. The concentration of samples should be high enough to be detected by the detector.

3. Preparative samples: In preparative separations, the size of the sample loop should be increased to increase the loading capacity. The technique using the dilute solution and repeatedly loading the samples should be applied instead of using a high concentration of samples.

(B) Buffer Preparation:

Salt concentrations and pH value are two factors that are greatly affected in the separation of biomolecules. A slight change of pH value of the buffer might affect the retention of biomolecules in the HYDROCELL column. The pH value of buffers needs to be adjusted to fit the specific separations in method development.

(C) Equilibrium Time

HYDROCELL columns are very hydrophilic. If the equilibrium time of the first buffer is not long enough, the samples may not have retention or the retention time are too short in the separation of biomolecules. The equilibrium volume must be at least 10 times column volume.

For example, a HYDROCELL column, 150 x 4.6 mm, has a column volume ( x 15) = 2.45 ml. The equilibrium volume of the column needs at least 24.5 ml. If the column is equilibrated at 1 ml / minute, it takes 24.5 minutes to equilibrate the column.

However, HYDROCELL columns are produced from rigid polymeric supports and have their mechanical strength. These columns can be equilibrated at high flow rates. The maximum flow rate for a 150 x 4.6 mm analytical column is 6 ml / minute. If the column is equilibrated at 3 ml / minute, it will take 8.16 minutes to equilibrate the column.

Back To Top


3. High Back Pressure

When HYDROCELL columns show back pressure higher than 1000 psi in the normal operation of flow rate, the columns are clogged. The columns need to be cleaned up using the procedure of the cleaning protocol or eluting with 1.0 M NaOH aqueous solution. After column cleaning, if the problems still persist, then the inlet frit of the column might be clogged and probably needs to be replaced. We always recommend the use of guard columns or precolumn filters to protect both analytical and preparative columns.

Back To Top

· Please call 812-234-2558 for additional technical support ·

 

Home | What's New | HYDROCELL HPLC Column | Biocompatible-Coating
Career Opportunities | Distributor Information | Ordering Information | Contact Us